h1. Spectrometer 

 h2. Functional Requirements (FR) 

 1. Measure light intensity at a fixed frequency (visible light) 
 > (a) Split light 
 > (b) Select frequency 
 > > i. Resolution 
 > > > A. Wavelength selectable in 1-10 nm steps 
 > (c) Measure the light intensity over the full spectrum of VIS light 
 > > i. Accurate -> no more than 1% off (5 nm) 
 > > ii. Reasonably fast -> each scan of the spectrum should take no more then XX second(s) 
 > > iii. Re calibration after scan should not be required. 
 > (d) Repeatable Measurements 
 > > i. Light intensity measured for a given sample at a fixed frequency should have minimal drift (< 1% ?) over the course of one experiment (2 hours) 

 2. Communicate data to student for plotting 
 > (a) Store data 
 > (b) Display data 

 3. Set up by instructor in < 15 minutes 

 4. Able to manufacture locally 

 5. User friendly for 1st year students 
 > (a) Easy to calibrate 
 > > i. Calibrating the light sensor takes less than 1 min 
 > > ii. Zero the spectrometer 
 > > iii. No additional equipment required for calibration 
 > (b) Independent of computer, or plug & play 

 6. Expected lifetime 3-5 years 
 > (a) Rugged 
 > > i. No irreversible damage from 15 cm drop to lab bench 
 > > ii. Chemical spills 
 > > > A. Conductive liquids do not damage electronics 


 * blah  

 h2. Design Parameters (DP) 

 * blah 

 h2. Vendors